5 | Plumbing and Priming the NGC Chromatography Systems
142 | NGC Chromatography Systems and ChromLab Software
3. Click Change Position to set the sample inject valve to Sample Pump Waste
and close its dialog box.
4. Click Start.
To purge the sample inlet valve (if available)
1. In the fluidic scheme, touch or click the Sample Pump module with the sample
inlet valve to open its dialog box.
2. In the Sample Pump dialog box, do one of the following:
If only one sample inlet valve is present, select S1 Port 8 as the port to
purge.
If two sample inlet valves are present, click Valve 2 and select S2 Port 8 as
the port to purge.