Bio-rad ChromLab™ Software Manuel d'utilisateur Page 105

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 178
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 104
Plumbing NGC Systems
Installation Guide | 103
8. Attach a 30 ml syringe to the 1/4-28 female to male luer adaptor.
9. Withdraw the plunger on the syringe until liquid is visible in the syringe.
10. Detach the luer adaptor and syringe from the pump head wash priming tubing
on pump B.
11. Locate the second inlet tubing with the 1/4-28 union that you set aside in
step 7.
12. Connect the union to the pump head wash priming tubing on system pump B.
13. Immerse the other end of the inlet tubing in the tube of 20% IPA.
To plumb the sample pump
Perform the steps in the section To plumb and prime the system pump wash lines
on page 102 with the following exception:
Attach the inlet tubing to the left rear port on the sample pump, and the pump
head wash priming tubing to the right rear port.
Vue de la page 104
1 2 ... 100 101 102 103 104 105 106 107 108 109 110 ... 177 178

Commentaires sur ces manuels

Pas de commentaire