12. Click OK. The spectrum flag changes from yellow to green for all 8 spectra, once calibration
is complete.
13. Export the data following the instructions in Appendix C and selecting the parameters for
export in the following order:
Array bar code
Spot name
Substance mass
Intensity
Resolution
S/N
Peak #
14. Paste the data into the ProteinChip SELDI OQ form, data sheet “Test 4 Mass Calibration”.
Paste the data into cell V41.
15. In the Internal Table, right-click on cell B21 and select Refresh Data.
16. The spreadsheet indicates if the test passed or failed. The test passes if the following
specifications are achieved:
Peptide average mass within 0.01% of calibrant mass
% Standard deviation (pooled CV) of <0.01
17. If the test fails, run maintenance procedure 1 (HV conditioning) and repeat the test. Contact
technical support if the test fails a second time.
28
10010680A.qxp 8/17/2007 1:59 PM Page 28