Bio-rad ProteinChip Qualification and Calibration Kits Manuel d'utilisateur Page 19

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 35
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 18
6. Run the same protocol on partition “n” of 4 for all 8 spots. Record the
partition number on the usage chart provided with the product packaging.
7. Make sure that a peak intensity of 50–500 µA is visible on each spot. If peak intensities
appear out of this range, change the laser energy and run the test again.
16
10010680A.qxp 8/17/2007 1:58 PM Page 16
Vue de la page 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 34 35

Commentaires sur ces manuels

Pas de commentaire