9. Plot all 8 spectra. Using the peak selection tool, select the insulin peak (5.963 kD) in all spectra.
10. Export the data following the instructions in Appendix C and selecting the parameters for
export in the following order:
Array bar code
Spot name
Substance mass
Intensity
Resolution
S/N
11. Paste the data into the ProteinChip SELDI OQ form, data sheet “Test 2 Mass Drift and
Resolution”.
12. The spreadsheet indicates whether the test passed or failed.
13. If the test fails, run the HV conditioning procedure (maintenance procedure 1) and repeat the
test. Contact technical support if the test fails a second time.