ProteinChip SELDI System: Reader Guide
10. Replace the load lock door by screwing it clockwise back on to the post.
11. Tighten the door until it seals against the O-ring.
12. Unscrew the load lock door about 1/2 turn until it is in the open position
(cutout in door should match opening in load lock).
13. Carefully replace the pawl drive assembly, lining up the posts on the load
lock door with the pawl drive plate. Take care not to catch the load lock
door home sensor wires between the pawl drive plate and the autoloader
mounting plate (Fig. 21).