ENrich
Q
ENrich
S
High-Resolution Ion
Exchange Columns
Instruction Manual
Catalog numbers
780-0001
780-0003
780-0021
780-0023
Please read these instructions before you use ENrich Q and S
high-resolution ion exchange media. If you have any questions or
comments regarding these instructions, please contact your
Bio-Rad Laboratories representative.
Vue de la page 0
1 2 3 4 5 6 ... 19 20

Résumé du contenu

Page 1 - Instruction Manual

ENrich™ QENrich™ SHigh-Resolution Ion Exchange ColumnsInstruction ManualCatalog numbers780-0001780-0003780-0021780-0023 Please read these instructions

Page 2

6 ENrich Instruction Manual Gradient Volumes and Salt ConcentrationsAs a starting point for developing a separation, we recommend using the ENrich c

Page 3 - Table of Contents

ENrich Instruction Manual 7Section 3: Care of the ENrich Ion Exchange Column 3.1 Column Cleaning Careful preparation (especially filtration) of th

Page 4

8 ENrich Instruction Manual 3.2 Bed Height Adjustment Under certain conditions of buffer composition, high flow rates or long-term use, the resin bed

Page 5 - ENrich Ion Exchange Columns

ENrich Instruction Manual 93.3 Frit ReplacementThe top frit may need to be replaced after extensive column use or if increasing backpressures are

Page 6 - Chromatography Systems

10 ENrich Instruction Manual 4. Raise the adaptor a few millimeters by slowly by turning the adjusting nut counterclockwise (Figure 5).Fig. 5. Rais

Page 7 - : Use of the ENrich

ENrich Instruction Manual 117. Stop the pump. Plug the bottom of the column. Set the column upright in a beaker. Remove the tubing and fitting fr

Page 8

12 ENrich Instruction Manual 13. Lower the adaptor to the top of the bed by turning the adjusting nut clockwise (Figure 8). Fig. 8. Lower the adapt

Page 9 - 2.4 Elution Conditions

ENrich Instruction Manual 13Section 4: Product Information Catalog Number Product Description780-0001 ENrich Q 5 x 50 Column780-0003 ENrich Q 10

Page 13 - 3.3 Frit Replacement

Life ScienceGroupSig 121310027324 Rev E US/EGBio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 424 6723 Australia 61 2 9914 2800 Austr

Page 14 - Fig. 6. Remove the retainer

Table of ContentsSection 1: Characteristics of the ENrich Ion Exchange (IEX) Columns ... 11.1 Introduction ...

Page 16 - 3.4 Storage Conditions

ENrich Instruction Manual 1Section 1: Characteristics of the ENrich Ion Exchange Columns 1.1 IntroductionENrich prepacked ion exchange columns me

Page 17 - : Product Information

2 ENrich Instruction Manual Table 1. Column characteristics.Q SColumn dimensions(diameter x height, mm)5 x 50 10 x 100 5 x 50 10 x 100Colu

Page 18

ENrich Instruction Manual 3Section 2: Use of the ENrich Columns2.1 Preparation for Initial Use The columns are supplied in a storage solvent of 2

Page 19

4 ENrich Instruction Manual Table 2. Buffers for ion exchange chromatography.Anion Exchange — ENrich QpH RangeBuffer MW pKa @ 25°CCounter-Ion pKa/°C

Page 20 - Laboratories, Inc

ENrich Instruction Manual 52.3 Sample Preparation Proper adjustment of the sample pH and ionic strength is critical for consistent and reproducib

Commentaires sur ces manuels

Pas de commentaire