Bio-Scale™ MT ColumnsInstruction ManualCatalog numbers751-0081751-0083751-0085751-0087
6 Bio-Scale MT Column Instruction Manual5. Attach the column to a pump and start flow at 50% of the desired operating flow rate.6. When all the air
Bio-Scale MT Column Instruction Manual 78. Pass 5 bed volumes of buffer through the column at a flow rate 20% higher than the operating flow rate. 9.
8 Bio-Scale MT Column Instruction ManualSection 3Care of the Bio-Scale MT Columns3.1 Frit RemovalThe top frit may need to be replaced after extens
Bio-Scale MT Column Instruction Manual 9Section 4Solvent Compatibility TableTable 2. Solvent Compatibility.SolventsAcetic acid (50%)AcetonitrileButa
10 Bio-Scale MT Column Instruction ManualSection 5Product InformationCatalogNumberProduct DescriptionBed Dimension, mmBed Volume, mlPressure Limit,
Bio-Scale MT Column Instruction Manual 11
Life ScienceGroupSig 12134006044 Rev B US/EGBio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 424 6723 Australia 61 2 9914 2800 Austri
Table of ContentsSection 1 Characteristics of Bio-Scale MT2, MT5, MT10, and MT20 Columns ... 11.1 Introduction ...
Bio-Scale MT Column Instruction Manual 1Section 1 Characteristics of the Bio-Scale MT2, MT5, MT10, and MT20 Columns1.1 IntroductionThe Bio-Scale MT
2 Bio-Scale MT Column Instruction Manual1.3 Column SpecificationsAll wetted parts are bio-compatible (Tefzel, polypropylene, polyphenyl sulfide, bor
Bio-Scale MT Column Instruction Manual 3Section 2Use of Bio-Scale MT ColumnsFig. 1. Diagram of Bio-Scale MT Column.FLOWUpper end fitting (1)Upper end
4 Bio-Scale MT Column Instruction Manual2.1 Column Disassembly (part given are in parenthesis)1. Unscrew and remove the upper (1) and lower (11)
Bio-Scale MT Column Instruction Manual 59. Gently turn the lock nut until the upper bed support is in the desired position.10. Attach upper (1) and