Bio-rad ENrich™ High-Resolution Size Exclusion Columns Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Instruments de mesure Bio-rad ENrich™ High-Resolution Size Exclusion Columns. Bio-Rad ENrich™ High-Resolution Size Exclusion Columns User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 16
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
ENrich
SEC 70
ENrich
SEC 650
High-Resolution
Size Exclusion Columns
Instruction Manual
Catalog numbers
780-1070
780-1650
Please read these instructions before you use ENrich SEC
high-resolution size exclusion media. If you have any questions or
comments regarding these instructions, please contact your
Bio-Rad Laboratories representative.
Vue de la page 0
1 2 3 4 5 6 ... 15 16

Résumé du contenu

Page 1 - Catalog numbers

ENrich™ SEC 70ENrich™ SEC 650High-Resolution Size Exclusion ColumnsInstruction ManualCatalog numbers780-1070780-1650 Please read these instructions be

Page 2

6 ENrich SEC Instruction Manual Section 3: Care of the ENrich SEC Column 3.1 Regular Maintenance To maintain performance, the column should be store

Page 3 - Table of Contents

ENrich SEC Instruction Manual 73.3 Column CleaningCareful preparation (especially filtration) of the sample and the buffers will maintain the colu

Page 4

8 ENrich SEC Instruction Manual Fig. 2. Column diagram.1. Start a slow flow rate of buffer (0.5 ml/min or less) through the column.2. Remove the lo

Page 5 - Size Exclusion Columns

ENrich SEC Instruction Manual 93.5 Storage Conditions To prevent bacterial growth in the column, it must be stored correctly. For short-term stor

Page 6 - Chromatography Systems

10 ENrich SEC Instruction Manual Section 4: Ordering Information Catalog # Description780-1070 ENrich SEC 70 10 x 300 Column, 10 x 300 mm780-1650 EN

Page 8 - 2.5 Column Calibration

Life ScienceGroupSig 121210027325 Rev A US/EGBio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 424 6723 Australia 61 2 9914 2800 Austri

Page 10 - : Care of the ENrich SEC

Table of ContentsSection 1: Characteristics of the ENrich Size Exclusion Columns ... 11.1 Introduction ...

Page 12 - Fig. 2. Column diagram

ENrich SEC Instruction Manual 1Section 1: Characteristics of the ENrich™ Size Exclusion Columns 1.1 IntroductionENrich prepacked columns for size

Page 13 - 3.5 Storage Conditions

2 ENrich SEC Instruction Manual The ENrich support is compatible with aqueous solutions of 6 M guanidine-HCl and 8 M urea. Detergents and organic so

Page 14 - : Ordering Information

ENrich SEC Instruction Manual 3Section 2: Use of the ENrich Columns2.1 Preparation for Initial Use The column is supplied in 20% ethanol. To remo

Page 15

4 ENrich SEC Instruction Manual 2.3 Sample considerations1. Always filter (0.22–0.45 μm filter) and/or centrifuge your sample to remove any particula

Page 16 - Life Science

ENrich SEC Instruction Manual 5Figure 1. Typical chromatograms for ENrich SEC media. Detection at 280 nm. Absorbance at 280 nm may be greater dep

Commentaires sur ces manuels

Pas de commentaire