Bio-rad ChromLab™ Software Manuel d'utilisateur Page 259

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 306
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 258
UV or UV/Vis Detector
Instrument Guide | 257
Sample trace values
are lower than
expected.
The lamps might be nearing the
end of their usable life.
Check the lamp usage values in
the System Information dialog
box. From the System Control
tab in ChromLab open File >
System Information > Detectors
tab.
If the lamp usage values are
close to their limits
(approximately 2,000 hours for
the deuterium and tungsten
lamps; approximately 5,000
hours for LEDs), change the
lamps.
For single-wavelength LEDs,
see Replacing the Single-
Wavelength UV Detector LED
on page 201.
For multi-wavelength lamps,
see Replacing the Multi-
Wavelength UV/Vis Detector
Lamps on page 209.
Symptom Possible Cause Possible Solution
Vue de la page 258
1 2 ... 254 255 256 257 258 259 260 261 262 263 264 ... 305 306

Commentaires sur ces manuels

Pas de commentaire