A | Maintaining the Instrument
228 | NGC Chromatography Systems and ChromLab Software
The following image shows an empty bay.
5. Store the module and the screws in a safe place for future use.
Tip: If you received a replacement module from Bio-Rad, use the
packaging provided with your shipment to return the damaged module to
Bio-Rad.
6. (Optional) Convert the bay to single wide or double wide.
See Converting Bays to Fit Modules on page 231.
7. Locate the module that you plan to install and remove it from its packaging.
Vue de la page 229