Bio-rad ChromLab™ Software Manuel d'utilisateur Page 190

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 306
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 189
A | Maintaining the Instrument
188 | NGC Chromatography Systems and ChromLab Software
Disassembling the Mixer
Use this procedure to replace any of the components of the mixer.
Important: Before disassembling the mixer, flush any hazardous material from
the NGC system. Ensure that the pumps are not running and bleed residual
pressure from the system.
To disassemble the mixer
1. On the touch screen, select Shut Down on the dropdown menu to exit
ChromLab and shut down the NGC instrument.
2. Disconnect the tubing from the top, bottom, and sides of the mixer.
Vue de la page 189
1 2 ... 185 186 187 188 189 190 191 192 193 194 195 ... 305 306

Commentaires sur ces manuels

Pas de commentaire