Bio-rad Bio-Scale™ Mini Macro-Prep® High Q and Macro-Prep Manuel d'utilisateur Page 14

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 28
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 13
Section 3
Connecting to Other Liquid
Chromatography Systems
Bio-Scale Mini cartridges can be connected to any
liquid chromatography system, provided that the
maximum pressure limit (3 bar, 45 psi, or 300 kPa) of
the cartridges is not exceeded. It is recommended
that the system pressure limit be set according to
the cartridge pressure limit. Pressures in excess of
3 bar are usually caused by restrictions in tubing or
detector cells downstream from the cartridge. Bio-
Rad offers two fitting kits for easy connection of a
Bio-Scale Mini cartridge to a BioLogic DuoFlow,
HPLC, or FPLC-type system.
3.1 BioLogic DuoFlow Systems
The luer to BioLogic system fittings kit (catalog
#732-0113) includes 1/4-28 female to male luer
and 1/4-28 female to female luer to connect one
Bio-Scale Mini cartridge to the BioLogic DuoFlow
system, see Figure 4.
10
10006400B.qxp 7/10/2006 3:48 PM Page 10
Vue de la page 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 27 28

Commentaires sur ces manuels

Pas de commentaire