2. To maximize gradient accuracy and to apply
samples efficiently, install 1.6 mm ID tubing
from the pump to the MV-6 sample inject valve
(if available). If using the MV-6 sample inject
valve, turn the knob counterclockwise as far as
it will go so it will now correspond to the printed
diagram on the valve (see Figure 2).
Fig. 2. Connecting to an MV-6 valve.
TO
COLUMN
TO
BIOLOGIC LP
SYSTEM OR ECONO
GRADIENT PUMP
TO
BIOLOGIC LP
SYSTEM OR ECONO
GRADIENT PUMP
MV-6
INJECT
VALVE
SAMPLE LOOP
WASTE
INJECT
PORT
FILL
TO
COLUMN
MV-6
INJECT
VALVE
SAMPLE LOOP
WASTE
INJECT
PORT
FILL
"INJECT" POSITION"FILL" POSITION