3. Connect the inlet of the cartridge to the male
luer fitting on the MV-6 sample inject valve (see
Figure 2). If not using the MV-6 sample inject
valve, connect a barb to male luer fitting on the
1.6 mm ID tubing, then connect to the top of
the female luer on the Bio-Scale Mini cartridge.
For optimum performance, a cartridge should
be mounted vertically with the arrow on the
cartridge pointing downward (see Figure 3).
4. Connect the cartridge outlet to the 1.6 mm ID
tubing leading to the BioLogic LP system optics
module or to the Model EM-1 Econo
UV
monitor. It is recommended to use the shortest
length (approximately 10 cm) of 1.6 mm ID
tubing. Connect a barb to female luer to the
1.6 mm ID tubing, then connect to the bottom
of the male luer on the Bio-Scale Mini car-
tridge.
8
10006400B.qxp 7/10/2006 3:48 PM Page 8