System Settings
User Guide | 7 7
4. Choose the Device Input tab.
5. By default, the fields in this dialog box are disabled. Select the Enabled
checkbox for SIM 1 and provide the following details to enable connection to
the first detector:
a. Type a trace name for the detector in the Trace Name field.
Note: This field has a 20-character limit.
b. Range (Volt) is the output from the detector. Set the minimum and
maximum volt range.
c. Type a unit type for the trace in the Trace Unit field.
Note: This field has a 5-character limit.
d. Scale converts the output to the trace’s units. Set the minimum and
maximum unit range (y-axis scale).
Vue de la page 78