5 | Method Editor
146 | NGC Chromatography Systems and ChromLab Software
NGC Scout Systems
Buffer selection: Manually Prepared Buffer via Blending Valve
Components to clean
Sample loop
Buffer blending valve
pH flow cell
Wash solution is delivered through Q1 for the sample loop and pH flow cell.
Ports Q1 and Q4 are washed individually with 5 ml.
Buffer selection: Buffer Blending
Components to clean
Sample loop
Buffer blending valve
pH flow cell
Wash solution is delivered through Q1 for the sample loop and pH flow cell.
Ports Q1, Q2, Q3, and Q4 are washed individually with 5 ml.