CHEF Mapper Year 2000Compliance Upgrade KitCatalog Number170-3684For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORA
Table of ContentsPageSection 1 Replacement of the ROM Chip...1Section 2 Verification
The CHEF Mapper requires a ROM chip upgrade in order to work properly afterDecember 31, 1999. This instruction sheet describes how to perform the Year
24. Note the large card cage to your left, when observing the rear of the Mapper. Press into removeconnector
5. Grasp and pull the second board from the right (Analog Interface Board) straight out ofthe card cage. Lay the card (component side down) on a clean
Unplug these two connectors from the board. The lower (blue) one has two clips on the topand the bottom. The picture above shows the clip on the botto
Now, replace this board in the Mapper, following the procedures outlined above in reverseorder:10. Re-seat the cables on the Processor I/O Board. 11.
3. To check the date and time, press CLOCK READ. The date and time are displayed for 5seconds. The time is displayed as hours:minutes:seconds, and the