10.7 Decontamination
Instruction Manual | 109
Note: The user must be present to change the containers approximately 1 hour into the
procedure. After this point the system can be left overnight to complete the rinse and shutdown
if that option is chosen.
The system will soak and go through an extended warmup (Figure 138) before a system startup
is commenced (Figure 139).
Fig. 138. Decontamination window while soaking system in proper fluidics and warming up.
Fig. 137. Decontamination window when priming rinse filter and nozzle path.