2.2 Description of System ComponentsThe Bio-Plex 200 system is comprised of the following components:• Array reader — combines 2 lasers, fluidics, an

Page 3 - Table of Contents

Table 1. Array Reader Front and Side Panel Features.Feature DescriptionSample arm The sample arm transports the sample from the 96-wellmicrotiter pl

Page 4

Table 2. Array Reader Rear Panel Features.Feature DescriptionCommunications The DB9-PIN connector is used to connect the array port P1 reader to the

Page 5

2.2.2 Microplate PlatformThe microplate platform (Figures 4 and 5) allows the automated processing of samples from a96-well microplate. The features

Page 6 - General Information

2.2.3 High-Throughput Fluidics (HTF) The Bio-Plex HTF (high-throughput fluidics) is designed to automate the introduction of sheathfluid into the arr

Page 7

2.2.4 Computer and MonitorThe Bio-Plex 200 system may be supplied with a computer. If so, please transfer the computer’sregistration to your company’

Page 8

2.3 Recommended Additional Equipment Not ProvidedSurge protectorWe recommend the use of a 6-outlet surge protector, with a minimum surge current of 1

Page 9 - Introduction

4. The maximum distance between the computer and the microplate platform and array readershould be 1.5 m (5 ft), the length of the communications cabl

Page 10

• Communication cable• Sheath fluid bottle• Sheath fluid waste bottle• 1 sample needle (11.6 cm/4.6 in)1. Unpack all components, being careful not to

Page 11

Fig. 12. Completed connections of computer and monitor to the Bio-Plex suspension array system.3.5 Connecting the Sheath Fluid and Waste Containers1

Page 13 - Communications port

Fig. 13. Waste and sheath fluid bottle connections.Note: To set up the 20 L sheath fluid cube container for use, remove tape from box and lift offthe

Page 14 - Table 5. HTF Features

162. Insure that the array reader, microplate platform, and HTF unit are turned off.3. Insert the Bio-Plex Manager CD-ROM into the CD drive of the com

Page 15 - Fig. 9. Bio-Plex reservoir

Fig. 15. Sample needle assembly.4. Remove the sample needle by grasping the needle and gently pushing up.5. Insert the new needle into the sample arm

Page 16 - Installation

Fig. 16. Sample needle dialog.5. Select the plate type being used.6. Click Eject/Retract to eject the plate holder.7. Place the MCV plate on the micr

Page 17 - 3.4 Array Reader Setup

Fig. 17. Sample needle assembly.12. By holding onto the needle height adjustment thumbscrew on the needle arm, manuallymove the needle so that it jus

Page 18

2. Click the Eject icon. Insert the MCV plate into the microplate platform. Click Retract.3. Choose Instrument from the main menu bar. Choose Addition

Page 19

Fig. 19. Device status window.6. After the pressure has stabilized, record the air and sheath pressure (Figure 19, arrows).Air pressure: ________ psi

Page 20 - 3.7 Software Installation

Fig. 20. HTF setup.9. Make the following connections to connect the HTF to the array reader:• Connect the sheath fluid line (blue fitting) to the She

Page 21

Note: You may use the additional waste line provided with the system to drain to a largerwaste container. The large waste container waste must be po

Page 22

15. Because the flow rate in step 14 is adjusted by altering the system pressure, it is necessaryto repeat steps 3 to 5 of Section 3.10 (Resetting Ins

Page 23

Table of ContentsPageSection 1 General Information ...11.1 About This Manual ...

Page 24 - 3.9 Initial System Priming

1 Make sure that the system has been set up according to the directions in the Millipore userguide.Hint: Make sure that the connector leading to the

Page 25 - 3.11 HTF Setup

3. Turn on the laboratory vacuum to maximum level (the same level that was used in the preceeding calibration procedure).4. Press on the filter plate

Page 26

Section 4Care and MaintenanceTable 6. Summary of Care and Maintenance.Daily Startup, calibrate, wash between plates, shut downWeekly Sonicate needle,

Page 27 - Fig. 20. HTF setup

Every MonthClean exterior surfaces Disconnect the instrument from AC power by turningoff the power switch on the rear of the array readerand micropla

Page 28 - Regulator

Air intake filter Replace the air intake filter every year. Disconnect the instrument from AC power by turning off the powerswitch on the rear of the

Page 29 - 3.12 Vacuum Manifold Setup

Section 5Troubleshooting5.1 Troubleshooting Guide for Bio-Plex®200 SystemProblem Possible Cause SolutionBio-Plex Manager™ Most Likely:has detected a

Page 30

Bio-Plex Manager Incorrect bead regions Compare bead regions in the assayhas detected a problem with selected in the protocol with those selected in

Page 31

Pressurizing in status bar of Leak in sheath bottle or Tighten sheath cap or replace sheathsoftware. cap bottle.Sheath and waste bottle Make sure that

Page 32 - Care and Maintenance

Bio-Plex Manager has Most likely:detected a change in sheath pressure. Sheath reservoir cap not Tighten sheath cap. Click OK. on securely Message sh

Page 33 - Every Year

Blue power indicator light on 1. System not plugged in Turn on system, power strip.HTF is not on 2. Power strip not on Remove power cable and reconn

Page 35 - Troubleshooting

5.2 Troubleshooting Guide for Vacuum ManifoldMessage/Problem Causes SolutionNo Flow/no vacuum. Lid on plate Remove lid.All wells not wet, or Wet unus

Page 36

5.3 Technical SupportFor technical assistance with the Bio-Plex 200 system, including all hardware and software,contact your local Bio-Rad office or,

Page 37

Microplate Platform SpecificationsInput voltage range 100–240 V, ~2.25 A, 47–63 HzPhysical dimensions (W x D x H) 44 x 61 x 8 cm (17.3 x 24 x 3 in)Wei

Page 38

Section 7Warranty StatementThis warranty statement may vary outside of the continental United States. Please contact yourlocal Bio-Rad office for the

Page 39

Section 8Ordering Information — System AccessoriesCatalog # DescriptionGeneral System Accessories171-000050 HTF171-000055 Sheath Fluid, 20 L171-0020

Page 40

CatalogNumber Product Description171-002030 Protective Shield for Sample NeedlePreventative Maintenance Items171-002032 Air Intake Filter(accessed t

Page 41 - Section 6

CatalogNumber Product Description171-203032 MCV plate IV, for use with Bio-Plex Manager 5.0and 4.1 Software171-203050 Bio-Plex Reservoir, for use with

Page 42 - Computer Specifications

5. Drain the sheath fluid and waste containers.6. Rinse the waste container with 10% household bleach solution and drain.7. Remove all specimens, d

Page 43 - Warranty Statement

Equipment Return Information Label andDECONTAMINATION CERTIFICATEBefore Bio-Rad can accept this equipment, you must certify that it is NOT CONTAMINATE

Page 46 - Decontamination Information

Life Science Group Sig 1109Bio-Rad Laboratories, Inc. Web site www.bio-rad.com USA 800 424 6723 Australia 61 2 9914 2800 Austria 01 877 89 01

Page 47 - Legal Notices

Section 1General Information1.1 About This ManualA Bio-Rad service engineer will install the Bio-Plex®200 system. However, the procedure is provided

Page 48

21.2.1 Electrical Safety InformationWarning: This instrument must be connected to an approved power source.Warning: Do not perform any maintenance or

Page 49

Caution: Removal of the array reader cover is intended for trained service personnel only. Donot attempt to operate the instrument with the cover remo

Page 50 - Laboratories, Inc

1.2.3 Mechanical Safety InformationCaution: During operation, this system contains exposed, moving parts. Risk of personal injuryis present. Keep han