Error Messages
Check Remedy
DIOM
Message Meaning Possible cause
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective
Replace the solenoid valve
16077 DIOM DIG12
current
16078 DIOM PWM9
status
Error: Status
Error: Over-current
Coil for solenoid valve defective
Replace the solenoid valve
Internal error DIOM Replace DIOM
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug.
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug.
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
16079 DIOM PWM10
status
Error: Status
16080 DIOM PWM11
status
Error: Status